In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address…
icscert·CWE-359·Published 2025-01-30
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
En su configuración predeterminada, el monitor de pacientes Contec Health CMS8000 transmite datos del paciente en texto plano a una dirección IP pública predefinida cuando el paciente se conecta al monitor. Esto podría provocar una filtración de datos confidenciales del paciente a cualquier dispositivo con esa dirección IP o a un atacante en un escenario de "máquina en el medio".
| Version | Type | Source | Base | Exp | Impact | Vector |
|---|---|---|---|---|---|---|
| 3.1 | Primary | cve.org | 5.9 | — | — | CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N |
| 3.1 | Secondary | NVD | 5.9 | 2.2 | 3.6 | CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N |
| 4.0 | Primary | cve.org | 8.2 | — | — | CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N |
| 4.0 | Secondary | NVD | 8.2 | — | — | CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X |